FS-16-12 slide stainer

Overall number of stations

16

Including:

 

1 drying station (adjustable temperature from +40° C to ~ +60° C)

1 station with flow-through trough (tap water)

14 combined stations (for troughs with reagents or for racks with slides)

Number of slides in a rack:

12

Volume of reagent in a  trough:

100 ml

Control and monitoring

Color touch-screen

Max. number of staining programs:

32

Max. number of program steps:

30

Free choice for each technological program:

  • Launch interval for processing of racks with slides (“interval”);
  • configuration (arrangement of troughs with reagent and “parking” stations for slide racks);
  • number of troughs with reagent for longest procession.

Free choice for each step:

  • Reagent;
  • station; time (0 sec – 59 min 59 sec);
  • activation period (1 - 99 sec);
  • draining time (0 - 99 sec);
  • number of dipping (1 - 99).

Safety

  • Internal ventilation system with connection either to an external ventilation system or to a charcoal filter;
  • electronic sensor which stops the process if the lid is open;
  • overflow sensor in the flow-through trough;
  • water consumption control device with solenoid valve.

Dimensions ( L / W / H )

460 × 520 × 330 mm

Weight

19 kg

Power supply

230 V / 50 Hz / 400 VA


FS-16-25 slide stainer

Overall number of stations

16

Including:

 

1 drying station (adjustable temperature from +40° C to ~ +60° C)

1 station with flow-through trough (tap water)

1 loading / unloading / parking station 

13 combined stations (for troughs with reagents or for racks with slides)

Number of slides in a rack:

25

 

Volume of reagent in a  trough:

210 ml 

Control and monitoring

Color touch-screen

Max. number of staining programs:

32

Max. number of program steps:

30

Free choice for each technological program:

  • Launch interval for processing of racks with slides (“interval”);
  • configuration (arrangement of troughs with reagent and “parking” stations for slide racks);
  • number of troughs with reagent for longest procession.

Free choice for each step:

  • Reagent;
  • station;
  • time (0 sec – 59 min 59 sec);
  • activation period (1 - 99 sec);
  • draining time (0 - 99 sec);
  • number of dipping (1 - 99).

Safety

  • Internal ventilation system with connection either to an external ventilation system or to a charcoal filter;
  • electronic sensor which stops the process if the lid is open;
  • overflow sensor in the flow-through trough;
  • water consumption control device with solenoid valve.

Dimensions ( L / W / H )

600 × 540 × 400 mm

Weight

25 kg

Power supply

230 V / 50 Hz / 400 VA


FS-9-12 slide stainer

Overall number of stations

12

Including:

 

1 drying station (adjustable temperature from +40° C to ~ +60° C)

1 station with flow-through trough (tap water) 

10 combined stations (for troughs with reagents or for racks with slides)

Blotter 

1 (for blotting the bottom side of a rack and slides’ ends)

Number of slides in a rack:

25 

 

Volume of reagent in a  trough:

210 ml  

Control and monitoring

Color touch-screen

Max. number of staining programs:

32

Max. number of program steps:

30

Free choice for each technological program:

  • Launch interval for processing of racks with slides (“interval”); 
  • configuration (arrangement of troughs with reagent and “parking” stations for slide racks); 
  • number of troughs with reagent for longest procession.

Free choice for each step:

  • Reagent; 
  • station; 
  • time (0 sec – 59 min 59 sec); 
  • activation period (1 - 99 sec); 
  • draining time (0 - 99 sec); 
  • number ofdipping (1 - 99). 

Safety

  • Internal ventilation system with connection either to an external ventilation system or to a charcoal filter; 
  • electronic sensor which stops the process if the lid is open; 
  • overflow sensor in the flow-through trough; 
  • water consumption control device with solenoid valve.

Dimensions ( L / W / H )

650 × 540 × 330 mm

Weight

22 kg

Power supply 

230 V / 50 Hz / 400 VA


FS-9-25 slide stainer

Overall number of stations

9

Including:

 

1 drying station (adjustable temperature from +40° C to ~ +60° C)

1 station with flow-through trough (tap water) 

7 combined stations (for troughs with reagents or for racks with slides)

Number of slides in a rack:

25 

Volume of reagent in a  trough:

210 ml

Control and monitoring

Color touch-screen

Max. number of staining programs:

32

Max. number of program steps:

30

Free choice for each technological program:

  • Launch interval for processing of racks with slides (“interval”); 
  • configuration (arrangement of troughs with reagent and “parking” stations for slide racks); 
  • number of troughs with reagent for longest procession.

Free choice for each step:

  • Reagent; 
  • station; 
  • time (0 sec – 59 min 59 sec); 
  • activation period (1 - 99 sec); 
  • draining time (0 - 99 sec); 
  • number ofdipping (1 - 99). 

Safety

  • Internal ventilation system with connection either to an external ventilation system or to a charcoal filter; 
  • electronic sensor which stops the process if the lid is open; 
  • overflow sensor in the flow-through trough; 
  • water consumption control device with solenoid valve.

Dimensions ( L / W / H )

460 × 540 × 330 mm

Weight

18 kg

Power supply

230 V / 50 Hz / 400 VA


Recommended protocol of staining (PAP-DIFF modification of Papanicolaou technique)

REAGENT (STATION)

TIME

MMSS

MLT

REGIME

(I - immersion,

D - dipping)

AGITATION

DELAY (s)

1

TAP WATER

30

I

15

2

2

DIST.WATER or  ACETIC ACID 0,5%

5

I

0

2

3

HEMATOXYLIN

300

I

0

0

4

TAP WATER

200

I

15

2

5

BLOTTER

0

I

0

0

6

PAP-2 (OG)

15

I

0

4

7

BLOTTER

0

I

0

0

8

ALCOHOL-1

10

I

2

2

9

BLOTTER

0

I

0

0

10

PAP-3 (EA)

300

I

0

6

11

TAP WATER

5

I

0

2

12

BLOTTER

0

I

0

0

13

ALCOHOL-2

10

I

0

2

14

ALCOHOL-3

10

I

0

2

15

ALCOHOL-4

10

I

0

2

16

XYLENE-1

10

I

0

2

Total time ~ 13:15. Performance with automated slide stainer FASTAINER-12-25 is about 100 slides per hour.